AC Series Class II B2 Biological Safety Cabinet B2-X
Introduction:
The AC Series Class II B2 Biological Safety Cabinet is a high-performance containment unit designed for laboratories handling hazardous biological materials and chemical compounds. Unlike traditional Class II cabinets, the B2 model provides 100% total exhaust, ensuring that all potentially harmful air is completely expelled from the laboratory, making it ideal for applications requiring both biological and chemical safety.
Features:
1. Color key display: large screen, more clear and dynamic display of various values of the safety cabinet, and has the function of appointment timing.
2. Automatically adjust the wind speed.
3. With power-off memory function, the state before power-off will be restored after power-on, and there will be visual prompts and alarms.
4. The integrated design of the operation area makes it easy to disassemble and clean.
5. The cabinet is separated from the base. The height of the adjustable base can be customized.
6. Interlocking function: UV lamp and front window; UV lamp and fan, LED lamp; fan and front window.
7. It has the function of appointment timing, which can automatically set the time of starting and shutting down the safety cabinet and the running time of ultraviolet lamps, fans and sockets.
8. The front window closes the double trigger signal, so that the sterilization and disinfection function of the ultraviolet lamp is normally turned on.
9. The negative pressure air duct is equipped with a foreign matter filtering structure to prevent foreign matter such as paper scraps from entering the fan system and affecting the normal operation of the product.
10. The front window glass has a one-key lift function.
11. The cabinet is designed with a 10° inclination angle, which is in line with ergonomic principles, with a larger viewing angle, convenient operation and more humanization.
12. The one-piece hand rest has a large fat surface, which has a larger contact area with the arm and is more comfortable.
Technical Parameters:
Model
BSC-1100IIB2-X
BSC-1300IIB2-X
BSC-1500IIB2-X
BSC-1800IIB2-X
External Size(W*D*H)
1100*775*2250mm
1300*775*2250mm
1500*775*2250mm
1800*775*2250mm
Internal Size(W*D*H)
940*600*660mm
1150*600*660mm
1350*600*660mm
1625*600*660mm
Tested Opening
Safety height 200mm(8″)
Max Opening
480mm
Inflow Velocity
0.53±0.025m/s
Downflow Velocity
0.33±0.025m/s
ULPA Filter
Two, 99.9995% efficiency at0.12μm, filter life indicator
Front Window
Motorized. Two-layer laminated toughened glass≥6mm. Anti UV
Noise
≤65dB
UV Lamp
30W*1
30W*1
40W*1
40W*1
UV timer, UV life indicator, emission of 235.7 nanometers for most efficient decontamination
LED Lamp
12W*2
14W*2
16W*2
16W*2
Illumination
≥1000Lux
Consumption
1300W
1300W
1300W
1300W
Waterproof Socket
Two, total load of two sockets: 500W
Display
LCD display: exhaust filter and downflow filter pressure, filter and UV lamp working time, inflow and downflow velocity, filter life, humidity and temperature, system working time etc.
Control System
Microprocessor
Airflow System
0% air recirculation, 100% air exhaust
Visual and Audio Alarm
Abnormal airflow velocity, Filter replacement, Front window at unsafe height, High filter pressure alarm, abnormal power failure
Material
Work Zone: 304 stainless steel
Main Body: Cold-rolled steel with anti-bacteria powder coating