Auto ELISA Processor BIOBASE8001

BIOBASE8001

Auto ELISA Processor BIOBASE8001 is a fully automatic detection equipment that integrates sample addition, incubation, shaking, washing, and reading. It is widely used in hospital laboratories, disease prevention and control centers, blood centers, and blood banks, animal and plant inspection and quarantine institutions, scientific research institutions, university laboratories and other places.

Features:

1. 8 sampling channels, 16 incubation boards, 3 washers.

2. Adopt the principle of pressure type liquid detection.

3. With TIP detection, clot detection function.

4. High-precision pipetting, the lowest volume can reach 5μl.

5. Position sensing, real-time monitoring of gripper status.

6. Multi-spacing and multi-strength adjustable grippers can be adapted to grasp and place microplates and plate covers of different sizes.

7. Dual thread control, can run simultaneously with the sampling arm without interfering with each other.

Auto ELISA Processor BIOBASE8001
Auto ELISA Processor BIOBASE8001

Technical Parameters:

Model BIOBASE8001
Sample&Reagent Unit Sample Racks 12*20 push-pull sample rack with detection
Sampling Channel 8 channels
Liquid Level Detection Air pressure level detection, needle blocking intelligent detection
TIP 5 TIP racks
Pipetting 5~1000μl, capable of parallel sampling
Pipetting Precision 100μl: CV≤1.0%, accuracy≤2.5%

1000μl: CV≤0.5%, accuracy≤1.0%

Reagent Position 6 reagent positions, 40 quality control positions +12 reagent bottles +25 reagent tanks(reagent tank with code identification)
Washing Unit Washer 3 units with on plate detection function
Washing Head 16-channel(32-pin), double row cleaning, hanging installation, with leakage recovery
Injection Uniformity CV<1.5%
Washing Residual <1μl(residual volume per well)
Reading Unit Reading Channel 8 independent photometric channels
Linear Range 0.000~3.000Abs
Repeatability CV≤1.0%
Stability ≤±0.003Abs
Optical Filters 4 standard filters(405, 450, 492, 630nm)

Total 10 filters positions

Reading Accuracy ≤±0.02Abs from 0.0 to 1.0 OD,

≤±0.03Abs from 1.0 to 2.0 OD

Sensitivity ≥0.01mg/L
Channel Difference ≤0.02Abs
Incubation Unit Incubators 16 shaking incubators
Cover Positions 4(16 covers)
Heating Method Dry heating with cover
Shaking Function Horizontal circular oscillation, 1~3 levels adjustable, amplitude 2.5mm
Temperature Range RT~60 °C
Temperature Uniformity ±0.5℃
Temperature Fluctuation ±0.5℃
Data Management Software Above Windows 7 system
LIS System Bi-direction, support HL7 protocol
Results Absorbance and results reviewable by software
Printer External optional
Built-in Scanner Optional
Other Function Biosafety Protection Optional(Negative pressure working zone, HEPA filter)
Disinfection UV lamp
Lighting LED
IAP Function Firmware can be upgraded online at any time
Working Condition Power Supply AC100~240V, 50/60Hz
Temperature 10~30℃
Humidity 40~70%
Size&Weight External Size(W*D*H) Machine: 1925*824*1652mm

Base Stand: 1925*792*777mm

Package Size(W*D*H) Machine: 2095*950*1160mm

Base stand: 2095*950*1005mm

Accessories: 1560*930*530mm

Net Weight Machine: 320kg

Base stand: 153kg

Accessories: 68kg

Gross Weight Machine: 420kg

Base stand: 245kg

Accessories: 122kg