Flexible Combination Mode
Dual-module
Single Chemistry Module + Single Immunoassay Module
Chemistry: 2000 Tests/hour; Immunoassay: 600 Tests/hour; Electrolytes: 600 Tests/hour
Triple-module
Dual Chemistry Modules + Single Immunoassay Module
Chemistry: 4000 Tests/hour; Immunoassay: 600 Tests/hour; Electrolytes: 1200 Tests/hour
Single Chemistry Module + Dual Immunoassay Modules
Chemistry: 2000 Tests/hour; Immunoassay: 1200 Tests/hour; Electrolytes: 600 Tests/hour
Quadruple-module Dual Chemistry Modules + Dual Immunoassay Modules Chemistry: 4000 Tests/hour; Immunoassay: 1200 Tests/hour; Electrolytes: 1200 Tests/hourTriple Chemistry Modules + Single Immunoassay Module Chemistry: 6000 Tests/hour; Immunoassay: 600 Tests/hour; Electrolytes: 1800 Tests/hourSingle Chemistry Module + Triple Immunoassay Modules Chemistry: 2000 Tests/hour; Immunoassay: 1800 Tests/hour; Electrolytes: 600 Tests/hour
Features :
01/ Intelligent System Software
Automatic retesting of abnormal samples
Intelligent sample scheduling
Automatic alarm handling
Automatic cleaning and maintenance
Reagent remaining level warning
02/Sample Processing System
Automatic sample rack type scanning
Sample quality identification (Optional)
Sample tube type identification
Maximum loading of 350 samples per batch
Emergency priority function
03/Immunoassay System
Tube-type magnetic microparticle enzymatic chemiluminescence
Non-contact mixing
Reagent and consumable loading and unloading without stopping the system
Single unit throughput 600 Tests/hour
04/Chemistry System
Reagent loading and unloading without system shutdown
Total reaction volume as low as 80μL
Single unit throughput 2000 Tests/hour
ISE module throughput 600 Tests/hour (Optional)
Emergency results reporting time as fast as 10min
Technical Parameters :
Model
MT8600
Chemistry Analyzer Module
Throughput
2000 Tests/hour; Optional ISE module 600 Tests/hour
Sample Type
Serum, plasma, whole blood, urine, cerebrospinal fluid, semen, stool
Loading Mode
Track loading, with emergency track
Sample Position
Maximum 350 samples loaded at one time, original tube loading with sample rack
Reagent Position
250 positions, 2 cold storage functional reagent disks
Sample Volume
1~35μL, increasing by 0.1μL
Reagent Volume
1~350μL
Reaction Volume
80~350μL
Probe
Liquid level sensing, random tracking, three-dimensional anti-collision protection
Wavelength Range
340nm~850nm
Reaction Temperature
Circulating water bath, constant temperature 37±0.1℃
External Dimensions(W*D*H)
1505*1225*1245mm
Net Weight
500kg
Immunoassay Module
Throughput
600 Tests/hour
Reaction Principle
Tubular magnetic particle enzymatic chemiluminescence
Sample Type
Serum, plasma, urine
Cleaning Module
Double cleaning module, triple magnetic separation cleaning technology, non-contact type mixing
Reagent Position
Double reagent disks, 50 reagent positions, 2~8℃ cold storage
Reaction Tube Capacity
2000 reaction tubes, support random loading
Incubation Position
A total of 238 incubation positions, 37±0.1℃, dry bath heating
Sample Position
Maximum 350 samples loaded at one time, original tube loading with sample rack
Other Functions
Reagent radio frequency identification, sample needle collision prevention, liquid level detection, automatic dilution, optional clot detection, etc.
External Dimensions(W*D*H)
1500*1225*1245mm
Net Weight
500kg
Sample Loading Module
Sample Position
350 sample positions, 450 samples delivered per hour
STAT Function
Equipped with emergency priority function
External Dimensions(W*D*H)
972*1225*1495mm
Net Weight
160kg
Others
Power Supply
AC220 50/60Hz (Standard); AC110 50/60Hz (Optional), 4950VA
Packing Size(W*D*H)
1685*1410*1460mm; 1685*1410*1460mm; 1410*1350*1170mm
Gross Weight
670kg; 650kg; 360kg