Features:
1. Throughput 60t/h.
2. 14 inch touch screen.
3. Tri-angle laser scatter + flow cytometry method + impedance method for RBC and PLT counting.
4. 3D holographic scattergram displays the accurate 5 part differentiation of WBC.
5. Large storage capacity: 100,000 results(including histogram, scattergram, patient information).
Technical Parameters:
Model |
BK-6310 |
Throughput |
60 tests/hour |
Assay Items |
5 parts, 29 parameters, 3 histograms, 3D scattergram |
Principle |
Tri-angle laser scatter, Flow cytometrymethod, 3D scattergram analysis, Impedance method for RBC and PLT counting, Cyanide-free method for HGB test |
Test Mode |
CBC mode, CBC+DIFF mode venous whole blood, Capillary whole blood and prediluted |
Parameters |
WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, MPV, PCT, PDW, P-LCR, P-LCC, NEU%, LYM%, MON%, EOS%, BAS%, NEU#, LYM#, MON#, EOS#, BAS#
4 research parameter: ALY%, ALY#, IG%, IG# |
Performance |
Item |
Linearity Range |
Carry Over |
CV |
WBC |
0~300*10^9/L |
≤0.5% |
≤2.0% |
RBC |
0~8.00*10^12/L |
≤0.5% |
≤1.5% |
HGB |
0~250g/L |
≤0.5% |
≤1.5% |
PLT |
0~3000*10^9/L |
≤1.0% |
≤4.0% |
Sample Volume |
CBC+DIFF mode: ≤20μl
CBC mode: ≤10μl |
Storage |
10,000 results including histogram, scattergram and patient information |
Interface |
4 USB ports, 1 LAN port Bi-direction LIS, support HL7 protocol, Internal RFID reader |
Power Supply |
AC220V±10%, 50/60Hz; 110V±10%, 60Hz; 400VA |
Package Size(W*D*H) |
575*490*690mm |
Gross Weight |
46kg |
Reagent Package Size |
330*330*330mm, 400*290*240mm, 400*290*240mm |
Reagent Gross Weight |
22kg, 2kg, 1kg |