- BIOBASE 2000
- BIOBASE
Features:
1. 1 robotic arm, 2 pipetting probes(10~1000μl).
2. 4 units 96 well microplates(independent incubating).
3. 1 unit reader&washer(auto reading and washing).
Technical Parameters:
Model | BIOBASE2000 | |
Sample Unit | Sample Racks | Standard 192 sample positions(expandable, up to 552) |
Pipetting | 10~1000μl | |
Dispensing System | 2 aspirating and dispensing probe( X-Y-Z movement) | |
Reagent Unit | Reagent Position | 23 reagent racks, editable |
Pipetting Range | 10~1000μl, 1μl stepping | |
Pipetting Precision | (100μl)CV≤0.5% | |
Washing Unit | Dispensing Time | 4 minutes to whole 96 to microplates |
Washing Probes | Two line 8-nozzle manifolds | |
Wash Containers | Cleaning fluid(15L), waste water(25L), Buffer 1(5L), Buffer 2(2L),
Buffer 3(2L), Buffer 4(2L), with liquid level-sensing, liquid shortage and full alarm |
|
Waste Containers | 25L with waste full sensor | |
Washing Residual | <2μl | |
Reading Unit | Reading Channels | 8 independent photometric channels, mono anddichromatic reading |
Absorbance | 0.0~3.0OD | |
Spectral Range | 400~700nm | |
Optical Filters | 2 standard filters(450, 630nm); 2 more on request (405nm/450nm/492nm/630nm) | |
Reading Accuracy | ±1% from 0~1.5OD, ±2% from 1.5~3.0OD | |
Incubation Unit | Incubators | 4 independent units |
Temperature Range | RT to 45℃ | |
Temperature Accuracy | ±0.5℃ | |
Data Management | Software System | Above Windows 7 system |
LIS System | Bi-direction, support HL7 protocol | |
Results | Absorbance and results reviewable by software | |
Printer | External optional | |
Working Conditions | Power Supply | AC220V±10%, 50/60Hz; 110V±10%, 60Hz, 400W |
Temperature | 10~30℃ | |
Humidity | 30%~80% | |
Size&Weight | External Size (W*D*H) | 1235*705*875mm |
Net Weight | 146kg | |
Package Size (W*D*H) | 1380*860*1435mm | |
Gross Weight | 220kg |