CK Series 1200-2500W Water bath

1. General This product is automatic setting of PID algorithm without overshoot, complete function, easy to use, is the best product, analog temperature control is indispensable for medical apparatus and instruments, laboratory instruments commonly used temperature control equipment. Extensively…

Category Tag

Characteristics

(1) With complete enclosed air- cooling compressor for refrigeration, low temperature thermostatic bath has outstanding advantages such as quick refrigeration and low noise.

(2) Refrigeration system is equipped with multi-purpose protection devices such as over heating, over current and so on.

(3) Temperature can be controlled by microcomputer, which can be operated simply, has alarming device for temperature under low temperature limit and over high temperature limit, and is equipped with PID automatic control.

(4) Use double windows in red and green, upper window displays measurement value in red, and lower window displays measurement value in green, both in LED.

(5) Intelligent microcomputer can adjust temperature setting allowance so that digital display distinguish-ability reaches 0.1℃.

(6) Special user PID can be adjusted.

(7) It has internal and external cycles, external cycle will discharge thermostatic liquid inside the bath, and establish No. 2 thermostatic bath, also it can discharge liquid outside as cooling or heating source, to lower (raise)temperature of external experimental vessels, and enlarge use scope.

Technical Parameters

Model CK-4005GD CK-4007GD CK-4010GD
Temperature Range -40–200℃ -40–200℃ -40–200℃
Volume 5L 7L 10L
Display Resolution 0.1 ℃ 0.1 ℃ 0.1 ℃
Pump Flow 20L/min 20L/min 20L/min
Heating power 800W 1200W 2500W
Cooling power 1322W 2329W 3610W
Drainage Yes Yes Yes
Voltage 220V/50Hz 220V/50Hz 220V/50Hz

 

CK-2003GD

CK-2003GD

CK-4005GD

CK-4005GD

Specific operation settings:

 Please add liquid medium (such as silicone oil) in the water tank. If the instrument is used above 80 °C, the addition liquid should have 2cm space under the groove to prevent the oil from expanding and overflowing at high temperature.

 Plug in the power cord and turn on the power switch on the back of the chassis.

 Turn on the red boat switch on the back of the device (this switch is the circulation pump control switch), adjust the flow adjustment knob under the switch to adjust the flow of the circulation pump.