Class III Biological Safety Cabinet

Introduction:

Class III Biosafety Cabinet is totally enclosed and gas-tight, it also can meet the requirement of operating Class I,II,III,IV pathogenic factor. It can be used in P3,P4 laboratory.

  • BSC-1100IIIX BSC-1500IIIX
  • BIOBASE

Features:

1. With memory function in case of power-failure.

2. UV lamp ensures work area sterilization effectively.

3. An angled cabinet front ensure an ergonomic working posture.

4. Exhaust air is double-filtered through high-quality ULPA filters with typical efficiency of 99.9995% for 0.12μm particles.

Technical Parameters:

Model BSC-1100IIIX BSC-1500IIIX
External Size(W*D*H) 1540*880*2000 mm 1790*880*2080 mm
Internal Size(W*D*H) 910*650*660 mm 1165*650*660 mm
Pass Box Size(W*D*H) Internal Size: 430*330*355mm
External Size: 575*425*495mm
Gloves One pair of 800mm butyl rubber gloves
Front Window 8 mm toughened glass, anti-ultraviolet radiation
Display LCD display
Pre-filter Polyester fiber, washable
Pressure Negative, ≥ 120 Pa
ULPA Filter Air Supply Filter: Efficiency 99.9995% at 0.12μm
First Exhaust Filter: Efficiency 99.9995% at 0.12μm
Second Exhaust Filter: Efficiency 99.9995% at 0.12μm
Noise ≤67dB
Alarm Audio and visual alarm, abnormal pressure, filter replacement
Work Zone Material 304 stainless steel
Illumination ≥1000lux
Consumption 1200W 1300W
LED Lamp 12W*2 16W*2
UV Lamp 18W*2, 8W*1 30W*2, 8W*1
Emission of 253.7 nanometers for most efficient decontamination
Power Supply AC220V±10%, 50/60 Hz; 110V±10%, 60Hz
Gross Weight 350kg 450kg
Package Size(W*D*H) 1635*1065*2160mm 1890*1065*2250mm
Standard Accessories Gloves, fluorescent lamp, UV lamp*3, remote control, pressure meter, drain valve
Scroll to Top