Introduction:
Infant Incubator is one of the important equipment used in medical institutions. It is mainly used for constant temperature cultivation, temperature resuscitation, oxygen delivery, rescue, hospitalization and observation of sick infants and premature infants. It is used in hospitals, confinement centers, neonatology, etc.

Category Tag

Features:

1. LCD display

2. Double control modes of bed temperature and skin temperature

3. Double-layer thermostatic cover, automatic air curtain device

4. Humidity control function

5. Air & Oxygen blender

6. Oxygen supply system

7. Baby cot is infinitely adjustable

8. 24 kinds of sound and light alarm, comprehensive protection of baby safety

Technical Parameters:

General Parameters Item Specifications
Power Supply AC220V, 50/60Hz(Standard) 1320VA; 110V 60Hz(Optional)
Size(L*W*H) 950*645*(1250-11450) mm
Maximum Heater Output Power(220VAC) Air chamber heater: 330W;

Water tank heater: 155W

Auxiliary Network Power Output AC220V/50Hz, both maximum permissible currents are 220V 1.5A
Heater Indication 0~100%, 10% increasing amount
Temperature Control Mode Bed temperature mode; The skin temperature mode
Bed Temperature Control Range 25.0~37.0℃; 37.1~39.0℃(Span mode)
Skin Temperature Control Range 34.0~37.0℃; 37.1~38.0℃(Span mode)
Skin Temperature Sensor Display Range 5.0~65.0℃
Heating Time ≤50min
Temperature Variability ≤0.5℃
Temperature Uniformity (Mattress in horizontal position) ≤0.8℃
Temperature Uniformity (Mattress in tilted position) ≤1.0℃
Difference Between Average Incubator Temperature And Control Temperature ≤1.5℃
Skin Temperature Sensor Accuracy ±0.3℃
Humidity Control The Running Time Of The Tank After It Is Filled With Distilled Water Humidity below 70%RH, at least 24h
Water Tank Capacity 1500ml
Humidity Display Range 0%RH~99%RH
Humidity Control Range CAH%RH~90%RH(1% per adjustment)
Humidity Control Accuracy ±10%
Oxygen Concentration Control Oxygen Concentration Display Range 0%~99%
Oxygen Concentration Display Resolution 1%
Oxygen Concentration Control Accuracy ±5% oxygen volume concentration
Oxygen Concentration Setting Range 21%~60% (1% per adjustment)
Oxygen Sensor Service Life The maximum use time is 10000h at 100% oxygen concentration
Transport Storage Temperature -20℃~+55℃
Atmospheric Pressure Range 500hPa~1060hPa
Relative Humidity ≤93%RH
Packing Size/ Gross Weight 1150*800*1600mm

190kg

The Work Environment Relative Humidity 30%RH~75%RH
Environment Temperature +20℃~+30℃
Pressure Range 700hPa~1060hPa
Service Altitude ≤2000m
Overvoltage Category
Pollution Level 2
Ambient Air Velocity ≤ 0.3m/s
Other Indicators Ambient Noise In The Infant Compartment Of The Incubator No more than 55dB(A) at stable temperature
Carbon Dioxide Concentration In The Infant Compartment A mixture of 4% carbon dioxide and air was fed into an 8mm diameter tube in the vertical direction from the mat to the top of the box at 750mL /min to a point 10cm above the mattress center and 10cm above the center. The concentration was less than 0 5% after stabilization
The Velocity Of Air Flow Over The Baby Mattress ≤0.35m/s
General

Alarm

Power failure alarm, fan alarm, sensor alarm, deviation alarm, over temperature alarm, water tank placement error alarm, water shortage alarm
System Alarm Mcu1 and Mcu2 communication error alarm, Mcu2 and Mcu3 communication error alarm, analog-to-digital conversion chip 2544 failure alarm, analog-to-digital conversion chip 2548 failure alarm, over temperature alarm, temperature heating system failure, humidity heating system failure

Configuration:

Model BK-3201(I) BK-3201(II)
Temp. Control Bed/Skin temperature Bed/Skin temperature
Display LCD LCD
Bed Temp. Control Range 25℃-39℃ 25℃-39℃
Skin Temp. Control Range 34℃-38℃ 34℃-38℃
Humidity Control Range CAH%~99%RH CAH%~99%RH
Oxygen Control Range 21%-99% /
Cabinet Yes Yes