Microwave Digestion System 12/18 Vessels

Model : MDS-12; MDS-18

Microwave Leakage : <5mw/cm2

Temperature Control Accuracy : ±0.3℃

Temperature Controlling Range : 50~400℃
Rotation : 360° Continuous Rotating
Size(W×D×H)(mm) : 640x630x690(Main device)

Microwave Digestion System Specifications:

Model MDS-12 MDS-18
Vessel Quantity 12 Vessels 18 Vessels
Vessel Volume 100 ml
Temperature Monitoring Contactless IR sensor
Temperature Control Scanning control of each vessel
Temperature Control Range 50~400℃
Temperature Control Accuracy ±0.3℃
Maximum Working Temperature 250℃
Inner Vessel Temperature Limit 300℃
Display 7 inches Touch Screen
Rotation Mode 360° Continuous Rotation
Vessel Material lnner: lmported TFM
Outer; Imported PEEK+Glass Fiber
Pressure Testing Contactless pressure sensor
Maximum Working Pressure 6 MPa
Pressure Control Range 0-10MPa
Pressure Control Accuracy 0.01 MPa
Power Supply AC 220V, 50/60Hz
Microwave Power 0-2000W Adjustable/ 0-3000W Adjustable
Microwave Frequency 2450Hz
Microwave Leak <5mw/cm2
Sizes(W×D×H)(mm) 640x630x690(Main device)
Shipping Size(W×D×H)(mm) 800x800x810(Main device); 620x590x560(Accessories)
N.W./G.W. (kg) 70kg(Main device) + 26kg (Accessories)
/120kg (Main device) + 31kg (Accessories)
75kg (Main device) + 33kg (Accessories)
/120kg (Main device) + 38kg (Accessories)

Microwave Digestion System Feature:

 

Special Designed Sample Digestion Vessel

The automatic vent and self-resealing structure which won national patent ensures the digestion vessels can automatically release pressure and instantly

reseal when a sudden over-pressure situation occurs. It prevents waste of batch samples and solutions and also avoids vessel damage.

Contact-less Temperature and Pressure Monitoring System

The advanced contact-less IR sensor could measure the real-time temperature of sample solutions in each digestion vessel. Meanwhile, the contact-less pressure sensor could monitor the real-time pressure of each vessel. This realizes isolation between temperature and pressure controlling system, also avoids sample cross-contamination. Each real-time temperature and pressure value is displayed during the whole digestion process, allowing a clear check of digestion conditions.

Large Storage Capacity

It can edit and store 255 types of programs according to user’s requirements and each method program can set parameters (temperature, pressure, time,microwave power) according to their own requirements which provides convenient operation for users.

Safety Protection System

With high-strength double locked security door, the real-time temperature and pressure monitoring, automatic adjustment of overpressure and over-temperature and abnormal sound monitoring, the device can run in a highly safe environment.

Power Adjustment

The variable frequency resonance or non-pulse(optional) continues rotating mode ensures high efficient and even microwave throughput. It can be set

between 0 ~2000W/0~3000W according to user’s requirements.

Microwave Digestion System Detail Show

Microwave Digestion System 12/18 Vessels
Microwave Digestion Vessel Structure
Microwave Digestion System 12/18 Vessels
MDS-12 Inner Structure
Microwave Digestion System 12/18 Vessels
MDS-18 Inner Structure