Sample Concentrator, Water Bath Type

Model : SC-WB1000

Sealing Temperature : RT.+5℃~99℃
Temperature Accuracy : ≤±1°C
Samples Position : 12 positions
Max Vertical Travel : 200mm
Size(W×D×H)(mm) : 390×300×850
The sample concentrator stands out for its precision sample concentration technology. The water bath design provides stable and uniform heating to ensure rapid evaporation of water from the sample, resulting in faster and more efficient sample concentration.

Category Tag

Sample Concentrator Specifications:

Model SC-WB1000
Temp. Range RT.+5℃~99℃
Time. Range 1min~99h59min
Display Accuracy ±0.1℃
Temp. Accuracy ≤±1℃
Temp. Uniformity (At 60℃) ≤±1℃
Heating Time (40℃-99℃) ≤30min
Samples Position 12 positions
Tube Size φ10~φ29 (Liquid volume 1ml~50ml)
Max Vertical Travel 200mm
Max Gas Pressure 200Kpa
Max. Gas Usage 15L/min
Gas Joint Diameter φ8mm
Consumption 1000W
Power Supply AC220V, 50/60Hz
Size(W×D×H)(mm) 390×300×850
Shipping Size(W×D×H)(mm) Main Machine: 500x470x340
Accessory: 830x340x360
N.W./G.W. (kg) Main Machine: 4.8/7
Accessory: 5.2/8

 

All components of the SC-WB1000 are made of corrosion-resistant, easy-to-clean, and durable materials, including the sample rack, support assembly, base, and gas dispensing system. The tubes are placed on the sample rack and support tray. The gas then passes through a flow meter to the dispensing system (the flow meter measures and controls the gas flow, ensuring that the gas is supplied within the appropriate range). Flexible tubing then directs the gas to the valve tube assembly at each location. Depending on the tube size and solvent volume, they can be individually raised or lowered to the correct height. A needle blows the gas to the surface of the solution, causing rapid evaporation of the solvent.

Sample Concentrator Feature:

  1. High-definition LCD display for easy operation and viewing of various setup parameters
  2. The water bath provides uniform heating, ensuring that the sample is subjected to even heat distribution throughout the heating process
  3. Gentle heat to heat the sample avoids damage to the sample from excessive temperatures and ensures sample integrity
  4. 12 sample positions for high productivity
  5. Each sample is blown independently of the other to avoid interference or contamination between samples.
  6. Equipped with liquid level sensor and dry burn alarm function to ensure safer operation process
  7. Lifting needle valve to control the air consumption of each position