Laboratory Multi-channel Mechanical Pipettes Feature
| Dispensing head rotates for effortless pipetting convenience |
| Individual piston and tip cone assembly allow for easy repair and maintenance |
| Compound material-made tip cone secures high sealing performance |
| Compatible with most universal tip brands |
Medical Multi-channel Pipettes Parameter
| 8-channel Adjustable Volume Pipettes | ||||||
| Volume Range | Increment | Test Volume(μl) | Accuracy error | Precision error | ||
| % | μl | % | μl | |||
| 0.5-10μl | 0.1μl | 10 | 1.50% | 0.15 | 1.50% | 0.15 |
| 5 | 2.50% | 0.125 | 2.50% | 0.125 | ||
| 1 | 4.00% | 0.04 | 4.00% | 0.04 | ||
| 5-50μl | 0.5μl | 50 | 1.00% | 0.5 | 0.50% | 0.25 |
| 25 | 1.50% | 0.375 | 1.00% | 0.25 | ||
| 5 | 3.00% | 0.15 | 2.00% | 0.1 | ||
| 50-300μl | 5μl | 300 | 0.70% | 2.1 | 0.25% | 0.75 |
| 150 | 1.00% | 1.5 | 0.50% | 0.75 | ||
| 50 | 1.50% | 0.75 | 0.80% | 0.4 | ||
| 12-channel Adjustable Volume Pipettes | ||||||
| Volume Range | Increment | Test Volume(μl) | Accuracy error | Precision error | ||
| 0.5-10μl | 0.1μl | 10 | 1.50% | 0.15 | 1.50% | 0.15 |
| 5 | 2.50% | 0.125 | 2.50% | 0.125 | ||
| 1 | 4.00% | 0.04 | 4.00% | 0.04 | ||
| 5-50μl | 0.5μl | 50 | 1.00% | 0.5 | 0.50% | 0.25 |
| 25 | 1.50% | 0.375 | 1.00% | 0.25 | ||
| 5 | 3.00% | 0.15 | 2.00% | 0.1 | ||
| 50-300μl | 5μl | 300 | 0.70% | 2.1 | 0.25% | 0.75 |
| 150 | 1.00% | 1.5 | 0.5% | 0.75 | ||
| 50 | 1.50% | 0.75 | 0.80% | 0.4 | ||
Precautions for using multi-channel pipettes
| Pipette guns that are sucking liquid should not be laid flat.
The liquid in the tip can easily contaminate the inside of the gun and may cause the spring of the gun to rust |
| The scale of the pipette gun should be adjusted to the maximum after each experiment, let the spring return to the prototype to extend the service life of the pipette gun |
| When sucking liquid, be sure to loosen your thumb slowly and steadily, and never loose it suddenly,
to prevent the solution from being sucked in too fast and rushing into the liquid extractor to corrode the plunger and cause air leakage. |